SIL rated oxygen measurement for % oxygen levels
PMA30
SIL rated
temperature controlled
flow alarm
Oxygen measurement in a 19" cassette.
The PMA30 provides fast, accurate and reliable oxygen measurement in a 19" cassette.
It has a SIL rating according EN61508 and includes an integral flow alarm.
The PMA30 can be ordered in two configurations:
- PMA30A/D has a combined analogue/digital meter
- PMA30/D has only a digital meter
Paramagnetism is significantly higher for oxygen than other common gases. The M&C magneto-dynamic measurement cell uses this physical phenomenon to provide the most accurate and reliable measurements available.
The small (2 ml) flow cell houses a robust sensor that can have a flow rate of up to 1 l/min. The small cell design improves speed of response and the paramagnetic sensor measures oxygen concentrations with a response time of less than 3 seconds.
The analyser is able to have 5 selectable ranges:
- 0 to 1%
- 0 to 3%
- 0 to 10%
- 0 to 30%
- and 0 to 100%
This allows this system to be used for trace level oxygen measurements as well as oxygen purity applications.
GENERAL |
||||
Measurement |
Oxygen |
|||
Configuration |
On-line |
|||
Measuring range |
Selectable for 0 to 1, 0 to 3, 0 to 30, 0 to 10, 0 to 30, and 0 to 100 vol% O2, linear. |
|||
Expanded measuring ranges |
Optional. Zero suppression for spreading of measuring range, adjusted by the standard ranges, for example 20 to 21 % (1 %), 18 to 21 % (3 %), 90 to 100 % (10 %), max. 97 to 100 %, incl. process pressure compensation type SDPD |
|||
External measuring range indication |
Potential free contacts for each selected range, contact rating max. 48 VDC, 0.5 A |
|||
Remote range control |
For all 5 ranges by means of external potential free contacts |
|||
Response time for 90 % FSD |
< 3 seconds at 60 Nl/hr air |
|||
Accuracy after calibration |
Deviation: analogue = ±1 % of span at range 3 to 100 % / digital indicator = 0.1 vol.% O2. ±2 % of span at range 1 % |
|||
Reproducibility |
Deviation: analogue = <1 % of span / digital = ±0.1 vol.% O2 |
|||
Influence of ambient temperature |
No influence up to +45 °C |
|||
Influence of barometric pressure |
The oxygen reading varies in direct proportion to changes of the barometric pressure |
|||
Influence of sample gas flow |
Variation of gas flow between 0 to 60 Nl/hr air will cause a difference of < 0.1 vol.% O2 |
|||
Sample gas inlet pressure |
0.01 up to 1 bar g, (PMA 30 required admission pressure for competent flow rate, no pump inside) |
|||
Sample gas outlet pressure |
Outlet of analyser must discharge freely into atmosphere. option: process pressure compensation PD in range of 0.4 to 1.6 bar absolute |
|||
Flow rate of sample gas |
Max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7 to 70 Nl/hr |
|||
Temperature of sample gas |
-10 to +40 °C, dry gas |
|||
O2 transducer temperature |
Fixed at +50 °C |
|||
Ambient temperature |
-10 to +45 °C |
|||
Storage temperature |
-20 to +60 °C, relative humidity 0 to 90 %rh |
|||
Certification |
||||
European |
CE marked |
|||
IP protection |
IP 20 EN 60529 |
|||
Electrical standard |
EN 61010 |
|||
Shipping |
||||
Shipping weight |
2.5 Kg |
|||
MECHANICAL |
||||
Dimension |
3U H x 28HP W x 230 mm L + approx. 60 mm installation space |
|||
Enclosure |
3U/28HP - 19-inch cassette European standard, silver |
|||
Wetted materials |
Platinum, Glass, Polypropylene, 316Ti Stainless Steel, FPM, Epoxy resin |
|||
ELECTRICAL / UTILITIES |
||||
Power |
Internal power unit 230 VAC standard, 115 VAC available ±15 %, 40 to 60 Hz. 35.5 VA |
|||
Connections |
Mains: 3-pole chassis plug. signal: 9 and 25-pole plug |
|||
Cable length |
2 m for mains supply |
|||
Sample gas connection |
1/8 " NPT i, optional hose connector DN 4/6 available |
|||
Outputs |
||||
Output signal |
0 to 10 VDC, non-isolated, load > 100 KΩ, for the range of 100 vol.%, and 0 to 20 mA or 4 to 20 mA for the chosen range, non-isolated, max load 300 Ω/900 Ω |
|||
Analogue/Digital indication |
Analogue: selectable for each range with a scale of 0 to 30 and 0 to 100 %. Digital: 3 ½ digit 9 mm high LCD for 0 to 100 % O2 reading, selectivity 0.1 vol.% O2 |
|||
All specifications subject to change |