Measurement of dissolved oxygen in ppb range
Digox 6.1 KS
high accuracy
short response time
automatic calibration
GENERAL |
||||
Measurement |
Trace oxygen |
|||
Measuring range |
0 to 1,000 μg/l O2, display up to 20,000 μg/l separated in 3 measuring ranges |
|||
Measurement error |
< 1 % related to the measured value or minimum ± 0.5 μg/l |
|||
Detection limit |
0.2 μg/l |
|||
Resolution |
0.01 μg/l, adjustable |
|||
Response time |
t90 < 8 s |
|||
Calibration |
Manual and automatic calibration with self test, possible in three areas |
|||
Data logger |
2,000 data sets in a continuous data logger |
|||
Flow |
3 to 20 l/h, compensated automatically optimal 10 l/h |
|||
Sample pressure |
0 to 8 bar, unpressurised outlet pressure peaks up to 16 bar |
|||
Sample temperature |
0 to 60 °C, automatically compensated |
|||
Minimum conductivity |
> 1.5 μS/cm |
|||
Ambient temperature |
0 to 40 °C |
|||
Signal outputs |
3 x active 0(4)…20 mA, freely selectable |
|||
Interface |
optional Ethernet, USB 2.0 |
|||
Certification |
||||
European |
CE marked. |
|||
Shipping |
||||
Net weight |
8.0 kg, mounted on stainless steel plate |
|||
MECHANICAL |
||||
Dimension |
580 x 335 x 140 mm (H x W x D) |
|||
Ingress protection |
IP-65 |
|||
ELECTRICAL / UTILITIES |
||||
Power |
100…240 VAC (50/60 Hz), 20 VA, optional 24 VDC (DG 75.10.00) |
|||
All specifications subject to change. |
DATASHEET: