A loop powered dew point transmitter for industrial gases
PLMa
4 to 20 mA loop powered
digital I/O for controlling relays
-110 °C to +20 °C dew point
A dew point meter for dry gases.
This dew point meter uses leading edge nanosensor technology to provide accurate and repeatable results from a compact (170 grams), loop powered, transmitter. With a range of -110 °C to +20 °C dew point (dp) it may be used in compressed air and a wide variety of industrial gases. This dew point meter is designed for harsh environments, and can be installed to monitor and control compressed air dryers, or ensure that high quality air is available for medical air, breathing air and instrument air applications where humidity levels are safety critical.
GENERAL |
||||
Measurement |
°C and °F dew point, ppmv, ppmw, µB H2O vapour pressure, grams of H2O/m3 and Lbs H2O/106 standard cubic feet in natural gas. |
|||
Measurement principle |
Nanosensor aluminium oxide |
|||
Configuration |
On-line |
|||
Range |
-110 to + 20 °C dew point |
|||
Temperature range |
-40 to + 70 °C |
|||
Accuracy |
±2 °C temperature corrected |
|||
Temperature accuracy |
±2 °C |
|||
Resolution |
< 0.01 °C |
|||
Repeatability |
0.8 °C |
|||
Sample flow |
> 1 LPM |
|||
Response time |
95 % of step change in 3 mins. |
|||
Process temp. range |
-20 to +60 °C |
|||
Ambient temp. range |
-40 to +85 °C |
|||
Certification |
||||
European |
CE marked |
|||
Calibration |
NIST traceable (NPL traceable option) |
|||
Shipping |
||||
Shipping weight |
0.17 Kg |
|||
MECHANICAL |
||||
Dimension |
82.6 L x 41.3 Dia mm (including connector) |
|||
Process threads |
3/4 "-16 UNF-2A and 1/4 " NPT |
|||
Enclosure material |
316 stainless steel IP67 |
|||
Material (wetted parts) |
PTFE |
|||
Pressure |
345 Bar (5000 PSIA) |
|||
ELECTRICAL / UTILITIES |
||||
Power voltage |
5 to 28 VDC two wire loop powered operation |
|||
Cable |
Two conductor cable min. 24 AWG (for total cable length > 1 mile min. 20 AWG). Must be shielded to meet CE requirements |
|||
Outputs |
||||
Current |
4 to 20 mA (housing isolated) with HART® |
|||
Digital I/O |
Standard |
|||
All specifications subject to change |