Humidity & temperature transmitter
HygroFlex5 - HF5
interchangeable HC2 probes
cable linked or direct mounting probes
digital and analogue outputs
The HygroFlex relative humidity (rh) and temperature measurement transmitter Model HF5 provides accurate and robust measurements for industrial applications.
The instrument can be wall-mounted, and provides analogue and digital outputs for a variety of humidity scales.
Probes may be mounted either directly on the instrument, for ambient condition monitoring, or via up to 100m of cable, for insertion into process environments that include high temperature or high pressure applications.
Any of the HC2 range of probes may be used, making the HF5 a flexible and user friendly device.
The HC2 range of humidity (H2O) and temperature measurement probes are a completely new type of probe, thanks to AirChip3000.
Rotronic have taken water vapour measurement to the next step, offering a system accuracy of ±0.8 %rh and ±0.1 K.
The AirChip3000 combines an ASIC, microcontroller and EPROM on one chip mounted within the probe, allowing total interchangeability of probes within a few seconds.
The HF5 provides humidity outputs in a choice of humidity scales:
- rh in %
- Temperature in °C or °F
- Absolute humidity in Kg/m3
- Dew point/frost point in °C or °F
- Wet-bulb temperature in °C
- Enthalpy in KJ/Kg
- Specific humidity in g/Kg
- Vapour concentration in g/m3
- Mixing ratio in g/Kg
- Saturation vapour pressure in hPa
- Partial water vapour pressure in hPa
- Water vapour saturation pressure in hPa
- Water vapour partial pressure
For further information about humidity measurement and hygrometry issues, please visit our Masterclass section.
GENERAL |
|||||
Measurement |
Humidity and temperature |
||||
Number of probes |
1 (any HC2 probe) |
||||
Sensor |
Compatible with all HC2 probes |
||||
Measurement range |
0 to 100 %RH |
||||
Accuracy at 23 °C ±5 K |
Probe dependent |
||||
Calculated parameters |
All psychrometric parameters |
||||
Max. cable length |
5 m or 100 m with AC3003 signal amplifier |
||||
Display |
Configurable graphic backlit LCD (Optional) |
||||
Firmware upgrade |
Via HW4 software |
||||
Certification |
|||||
CE / EMC compatibility |
EMC directive 2004 / 108 / EC |
||||
Fire protection class |
Corresponds to UL94-HB |
||||
FDA / GMP conformity |
Conforms to 21 CFR Part 11 and GAMPS |
||||
Housing material /protection |
ABS |
||||
IP rating |
IP65 (models with USB or ethernet interface, IP40) |
||||
Probe material |
Probe dependent |
||||
Probe connection /interface |
E2 (threaded coupling) / UART |
||||
HF520 |
HF53x |
HF54x |
HF56x |
HF55x, digital |
|
Environmental |
|||||
HF5 operational range (with LCD) |
-10 to 60 °C , 0 to 100 %RH |
||||
HF5 operational range (without LCD) |
-25 to 60°C |
-40 to 60°C |
|||
MECHANICAL |
|||||
Dimensions (duct & wall mount) |
129 x 72 x 45 mm |
192 x 102 x 52 mm |
129 x 72 x 45 mm |
||
Weight |
220 g |
500 g |
220 g |
||
ELECTRICAL AND UTILITIES |
|||||
Configuration |
2 x 2 wire |
3/4 wire |
|||
Electrical connections |
Screw terminals inside M16 cable gland Socket (USB/Ethernet) |
2xM16 cable gland |
Screw terminals inside M16 cable gland Socket (USB/Ethernet) |
||
Power supply |
10 to 28 VDC |
15 to 40 VDC 12 to 28 VDC galv. isolated 9 to 36 VDC 7 to 24 VAC |
85 to 240 VAC |
Power over Ethernet (PoE) IEEE 802.3 af |
|
Current consumption |
2 x 20 mA max. |
270 mA max. (without Ethernet), 420 mA max. (with Ethernet) |
30 mA max. (without Ethernet), 45 mA max. (with Ethernet) |
CLASS 1 |
|
Analogue output options |
|||||
Number |
2 |
2 |
2 |
2 |
No analog outputs |
Current |
4 to 20 mA |
0(4) to 20 mA |
|||
Voltage |
N/A |
0 to 1/5/10 V |
|||
Galvanic isolation |
N/A |
HF54 and HF56 |
|||
Maximum load |
2x500 Ω |
≤2x500 Ω (current output)≥1 kΩ/V (voltage output) |
|||
Digital output options |
|||||
RS-485 |
No digital outputs |
RS-485 & analog |
RS-485 |
||
USB |
USB & RS-485 & analog |
USB & RS-485 |
|||
Ethernet |
Ethernet RJ45 & RS-485 & analog |
Ethernet RJ45 & RS-485 |
|||
Wireless |
Wireless & RS-485 & analog |
Wireless & RS-485 |
|||
All specifications subject to change |