Coulometric titration system for water in oil
Aquamax KF Coulometric
simple operation
10 user programmable methods
1 ppm to 100 %
A Karl Fischer coulometric titration system.
The Aquamax KF Coulometric Titrator has been developed over many years with suggestions and comments from users worldwide. This laboratory based system can be used in the field and on offshore platforms and includes an integral high speed printer.
It is battery operated, with an optional carry case to provide a convenient and flexible measurement system. The Automatically Compensated Errors (ACE) program makes it ease of use and portability enables it to be used by both laboratory technicians and field engineers.
Specifically designed to determine the water (H2O) content of oil and hydrocarbon liquids by measuring the amount of electrolysis current necessary to produce the required iodine. This is an absolute technique which does not require routine calibration of reagents. With this simple-to-programme device, a titration measurement may be taken with a single button.
The system conforms to:
ASTM D1533, D4928, D6304, IP386, IP4138,
API MPMS Chapter 10.9
IEC60814, ISO 10101-3, 10337 and 102937
The Aquamax KF Coulometric has a small footprint and results may be downloaded via the Results Manager software onto a spreadsheet programme.
Results Manager
Results Manager is supplied as standard with the Aquamax KF Coulometric and is a Windows based application, allowing users to view and print the results. The system can either be connected directly to a computer or results may be stored on a flash drive for later connection to a computer.
GENERAL |
||||
Measurement |
||||
Titration method |
Coulometric Karl Fischer titration |
|||
Configuration |
Bench-top |
|||
Electrolysis control |
Patented "ACE" control system GB2370641 |
|||
End point detection |
AC polarisation |
|||
End point indication |
Visual display/print out/acoustic beep |
|||
Typical measuring range |
1 µg to 10 mg water |
|||
Possible measuring range |
1 µg to 200 mg water |
|||
Moisture range |
1 ppm to 100 % water |
|||
Max. sensitivity |
0.1 µg |
|||
Precision |
10 to 100 µg ± 3 µg 100 µg to 1 mg ± 3 µg Above 1 mg ± 0.3 % |
|||
Max. titration speed |
2.24 mg per minute |
|||
Start delay time |
0 to 30 minutes, user selectable |
|||
End delay time |
0 to 30 minutes, user selectable |
|||
Max current |
400 mA |
|||
Drift compensation |
Automatically controlled |
|||
Calculation modes |
Weight/weight (W/w), Weight/dilution ratio (W/K), Volume/density (V/SG), Volume/volume(V/v) |
|||
Stirrer speed |
Microprocessor controlled |
|||
Display |
40 character alphanumeric backlit LCD |
|||
Printer |
42 character high speed thermal printer |
|||
Data entry |
15 key touchpad |
|||
Display format |
µg, mg/kg, ppm, % |
|||
Print format |
µg, mg/kg, ppm, % |
|||
Statistics |
max, mean, min values up to 99 runs |
|||
Method storage |
10 user programmable methods |
|||
Sample ID number |
User programmable |
|||
Languages |
Multiple languages, user selectable |
|||
Calendar/clock |
Analysis time and date print out |
|||
Battery low indicator |
Display and print out indication |
|||
Certification |
||||
European |
CE marked |
|||
Shipping |
||||
Shipping weight |
3.5 Kg |
|||
MECHANICAL |
||||
Dimension |
250 x 245 x 120 mm |
|||
Titration vessel |
Low drift cell design. No grease or PTFE sleeves required |
|||
ELECTRICAL / UTILITIES |
||||
Power |
90 to 264 VAC, 47 to 63 Hz. 12 VDC car adapter/internal battery |
|||
Outputs |
||||
Data |
RS232 ports and USB |
|||
All specifications subject to change |
PUBLISHED ARTICLES:
Recent advances in coulometric titration for water content
MANUALS:
Instruction Manual for Aquamax KF Coulometric
MSDS:
61001 Cou-Lo Formula A Coulometric Anode Solution
61002 Cou-Lo Formula C Coulometric Cathode Solution
61003 Cou-Lo Formula Water Solutions
61004 Cou-Lo Formula Water Solutions
DATASHEET:
Datasheet for Aquamax KF Coulometric