Calibration system for rh and temperature sensors
HygroGen2
< 0.1% rh stability
cost effective calibration
calibration over 0 to 60°C
Stable humidity and temperature calibration.
The Hygrogen2 provides a stable environment in which to calibrate any rh & temperature measuring instrument. This system, also available to hire, is a cost effective calibration solution.
Reaching humidity equilibrium in around 5 minutes, the Hygrogen2 can calibrate up to 5 probes simultaneously. The user friendly touch screen has a graphical interface and is capable of setting up to 200 set-point changes per programme with up to 20 programs.
The provision of a heated sample loop for the connection of a chilled mirror instrument allows cross checking with the control HygroClip sensor, and improves uncertainty.
The internal water reservoir incorporates onscreen level monitoring so the user can monitor and control the water level for the duration of a calibration.
An integrated UV sterilisation system within the water reservoir eliminates any issues caused by water-borne contaminants.
The standard Hygrogen2 is supplied with:
- touch screen user interface
- set-point control and programmer function
- heated sample loop
- 1x desiccant cell
- 1x fill syringe
- embedded HW4-P software.
A variety of chamber doors (ordered separately) are available to suit sensors with different diameters.
|
Relative humidity |
Temperature |
||
GENERAL |
||||
Generation method |
Mixed flow with desiccant dryer cell and piezoelectric humidifier |
Peltier thermoelectric element with radial chamber mixing fan |
||
Range |
5 to 95 %rh |
0 to 60 °C |
||
Programmer function |
20 user program memory, up to 200 set-point changes per programme |
|||
Stability at equilibrium |
< 0.1 %rh |
< 0.02 °C |
||
Chamber temperature gradients |
< 0.05 °C (15 to 50 °C), < 0.1 °C (5 to 60 °C), ±0.15 at 0 °C |
|||
Sensor |
HygroClip2-S, capacitive RH sensor, Pt100 temperature sensor |
|||
Type |
Embedded multiple input PID controller, touch screen graphical user interface |
|||
Response time |
3 mins (35 to 80 %rh change) |
5 mins (20 to 30 °C change) |
||
Control probe specification |
±0.8 %rh (23 °C ±5) ±2 %rh (0 to 60 °C) |
±0.1 K (23 °C ±5) ±0.3 K (0 to 60 °C) |
||
Typical calibration uncertainty |
±1.5 %rh (k=2) at 23 °C |
±0.15 °C (k=2) 15 to 50 °C |
||
Water level |
Low and high alarm, bar graph status indication |
|||
Water quality |
UV sterilisation, auto time cycling |
|||
Desiccant condition |
Condition monitored during control operation |
|||
Certification |
||||
CE |
Safety: EN 61010-1:2001 EMC: EN 61326-1:2006 EN 61326-1:2006 & EN 61000-6-1:2007 |
|||
MECHANICAL |
||||
Reference connection |
Temperature controlled outlet and inlet, 6mm fittings |
|||
Chamber volume |
2 litres, effective working volume 1.5 litres |
|||
Enclosure |
Powdered coated aluminium/steel, IP20, 450 x 406 x 205 mm (Max.) |
|||
Weight |
13 Kg |
|||
ELECTRICAL |
||||
Power |
110 to 240 VAC 50/60 Hz, 3A |
|||
USB ports |
7 front, 2 rear |
|||
All specifications subject to change |
DOORS AND FITTINGS |
||||
HG2-D-11111HG2 door 5x 15 mm ports (for 5 HygroClips) with 5 bungs. Order specific B1 sleeves for smaller probe diameters |
||||
HG2-D-88888HG2 door 5x 30 mm ports with 5 bungs. Order specific B8 sleeves for smaller probe diameters |
||||
HG2-D-HFWHygroGen probe door for HF3 and HF4 wall mount transmitters |
||||
HG2-DP-00000HG2 clear acrylic door (no ports) for instruments with displays |
||||
HG2-D-xxxxxHG2 fully custom door for more than 300 mm, see xx diameter codes |
||||
|
HG2-B115 mm Bung |
|||
|
HG2-B1-xxB1 probe sleeves (15 mm external, internal probe diameter see xx codes) |
|||
HG2-B8300 mm Bung |
||||
HG2-B8-xxB8 probe sleeves (30 mm external, internal probe diameter see xx codes) |
||||
SPARES |
||||
HG2-DCAdditional desiccant cell, pre-filled with molecular sieve |
||||
HG2-FILLFill tube and syringe |
||||
HG"-DES-3Molecular sieve desiccant (3 kg) |
||||
ACCESSORIES |
||||
HG2-AC3001-L/050HygroClip2 calibration cable, 50 cm, HC2 to USB |
||||
HG2-AC3001-L/050(5)HygroClip2 calibration cable, 50 cm, HC2 to USB (5 unit set: HG2-AC3001-L/050) |
||||
HG2-TBTransit bag, lightweight |
||||
HG2-TCHygroGen transit case |
||||
AC3015Mini USB transfer cable, 30 cm, with 90° coupling for transmitter with fixed probe |
MANUALS:
Instruction manual for HygroGen2
CALCULATORS AND CONVERSION CHARTS:
TECHNICAL BULLETINS:
Explanation of relative humidity