Calibration system for rh and temperature sensors

HygroGen2

< 0.1% rh stability
cost effective calibration
calibration over 0 to 60°C

Download a datasheetRequest a quoteCall us

 

Stable humidity and temperature calibration.

The Hygrogen2 provides a stable environment in which to calibrate any rh & temperature measuring instrument. This system, also available to hire, is a cost effective calibration solution.

Reaching humidity equilibrium in around 5 minutes, the Hygrogen2 can calibrate up to 5 probes simultaneously. The user friendly touch screen has a graphical interface and is capable of setting up to 200 set-point changes per programme with up to 20 programs. 

The provision of a heated sample loop for the connection of a chilled mirror instrument allows cross checking with the control HygroClip sensor, and improves uncertainty.

The internal water reservoir incorporates onscreen level monitoring so the user can monitor and control the water level for the duration of a calibration.

An integrated UV sterilisation system within the water reservoir eliminates any issues caused by water-borne contaminants.

The standard Hygrogen2 is supplied with:

  • touch screen user interface
  • set-point control and programmer function
  • heated sample loop
  • 1x desiccant cell
  • 1x fill syringe
  • embedded HW4-P software.

A variety of chamber doors (ordered separately) are available to suit sensors with different diameters.

 

Relative humidity

Temperature

GENERAL

Generation method

Mixed flow with desiccant dryer cell and piezoelectric humidifier

Peltier thermoelectric element with radial chamber mixing fan

Range

5 to 95 %rh

0 to 60 °C

Programmer function

20 user program memory, up to 200 set-point changes per programme

Stability at equilibrium

< 0.1 %rh

< 0.02 °C

Chamber temperature gradients

< 0.05 °C (15 to 50 °C), < 0.1 °C (5 to 60 °C), ±0.15 at 0 °C

Sensor

HygroClip2-S, capacitive RH sensor, Pt100 temperature sensor

Type

Embedded multiple input PID controller, touch screen graphical user interface

Response time

3 mins (35 to 80 %rh change)

5 mins (20 to 30 °C change)

Control probe specification

±0.8 %rh (23 °C ±5) ±2 %rh (0 to 60 °C)

±0.1 K (23 °C ±5) ±0.3 K (0 to 60 °C)

Typical calibration uncertainty

±1.5 %rh (k=2) at 23 °C

±0.15 °C (k=2) 15 to 50 °C

Water level

Low and high alarm, bar graph status indication

Water quality

UV sterilisation, auto time cycling

Desiccant condition

Condition monitored during control operation

Certification

CE

 Safety: EN 61010-1:2001 EMC: EN 61326-1:2006 EN 61326-1:2006 & EN 61000-6-1:2007

MECHANICAL

Reference connection

Temperature controlled outlet and inlet, 6mm fittings

Chamber volume

2 litres, effective working volume 1.5 litres

Enclosure

Powdered coated aluminium/steel, IP20, 450 x 406 x 205 mm (Max.)

Weight

13 Kg

ELECTRICAL

Power

110 to 240 VAC 50/60 Hz, 3A

USB ports

7 front, 2 rear

All specifications subject to change

 

 

 

DOORS AND FITTINGS

HG2 door with 15 mm diameter inputs

HG2-D-11111

HG2 door 5x 15 mm ports (for 5 HygroClips) with 5 bungs. Order specific B1 sleeves for smaller probe diameters

 HG2 door with 30 mm diameter inputs

HG2-D-88888

HG2 door 5x 30 mm ports with 5 bungs. Order specific B8 sleeves for smaller probe diameters

 

HG2-D-HFW

HygroGen probe door for HF3 and HF4 wall mount transmitters

 

HG2-DP-00000

HG2 clear acrylic door (no ports) for instruments with displays

 

HG2-D-xxxxx

HG2 fully custom door for more than 300 mm, see xx diameter codes

 

HG2-B1

15 mm Bung

 Probe sleeve, outside diameter 15 mm HG2-B1

HG2-B1-xx

B1 probe sleeves (15 mm external, internal probe diameter see xx codes)

 

HG2-B8

300 mm Bung

 Probe sleeve, outside diameter 30 mm HG2-B1

HG2-B8-xx

B8 probe sleeves (30 mm external, internal probe diameter see xx codes)

SPARES

 

HG2-DC

Additional desiccant cell, pre-filled with molecular sieve

 

HG2-FILL

Fill tube and syringe

 

HG"-DES-3

Molecular sieve desiccant (3 kg)

ACCESSORIES

 

HG2-AC3001-L/050

HygroClip2 calibration cable, 50 cm, HC2 to USB

 

HG2-AC3001-L/050(5)

HygroClip2 calibration cable, 50 cm, HC2 to USB (5 unit set: HG2-AC3001-L/050)

 Soft bag for HygroGen2

HG2-TB

Transit bag, lightweight

 

HG2-TC

HygroGen transit case

 

AC3015

Mini USB transfer cable, 30 cm, with 90° coupling for transmitter with fixed probe