Humidity Standards for Calibration
EAxx-SCS
Applications
On-site calibration and adjustment of ROTRONIC probes (third-party probes also possible).
With the humidity standards, a calibration device and the HW4 software, this can be done easily at a PC. Without the software it is also possible to calibrate and adjust probes with the handheld instrument HP23-A.
The most frequently used rh levels for calibration are 10 %, 35 %, 80 %.
FEATURES |
|||||
|
Ampoules contain unsaturated salt solutions |
|||||
|
Inexpensive calibration on site |
|||||
|
Simple and safe use |
|||||
|
Unlimited lifetime |
|||||
|
Practical packs of 5 ampoules of the same humidity value (approx. 0.8 ml per ampoule) |
|||||
AVAILABLE STANDARDS |
|||||
Humidity Value |
Uncertainty at 23 °C |
Order Code |
|||
|
0.5 %rh |
±0.3 %rh |
EA00-SCS |
|||
|
5 %rh |
EA05-SCS |
||||
|
10 %rh |
EA10-SCS |
||||
|
11 %rh |
EA11-SCS |
||||
|
20 %rh |
EA20-SCS |
||||
|
35 %rh |
±0.5 %rh |
EA35-SCS |
|||
|
50 %rh |
±0.9 %rh |
EA50-SCS |
|||
|
60 %rh |
EA60-SCS |
||||
|
65 %rh |
EA65-SCS |
||||
|
75 %rh |
EA75-SCS |
||||
|
80 %rh |
±1.2 %rh |
EA80-SCS |
|||
|
95 %rh |
EA95-SCS |
||||
COMPATIBLE PRODUCTS |
|||||
|
All calibration devices |
|||||
INCLUDED |
|||||
|
SCS traceable calibration certificate |
|||||
|
Textile pads |
|||||
|
Calibration instructions |
|||||
TYPICAL ACCESSORIES |
|||||
|
Textile pads in tubes (50 pc) |
EA-PADS |
||||

