Oil in water analyser for direct insertion into process line
EX-100P-1000P
direct insertion probe
automatic ultrasonic cleaning
fast response
Improved oil in water measurements.
Significant steps have been taken to improve oil in water measurements. These systems from Advanced Sensors have been able to operate in hazardous areas for many years with zero maintenance.
Similar to the EX-100 and EX-1000, these systems use a laser induced UV fluorescence technique to monitor trace levels of oil in water. The EX-100P monitors the total fluorescence, while the EX-1000P provides an integral spectrometer to identify different oil contaminants or hydrocarbon liquids. Both systems can monitor from low ppb to 20,000 ppm, and are well suited to monitoring water pollution prior to discharge or re-injection. These versions of the analyser continuously measure a sample stream, with a probe inserted directly into a main process line. A variety of pressure rated flanges are available. The probe may be inserted directly into the process line without flow conditioning or flow control, and is recommended for applications where rapid oil trending information is required.
The use of an automatic ultrasonic cleaning method allows the EX-100P to provide reliable and accurate measurements for many years without requiring maintenance or consumables, making the operating costs for these systems extremely low. Many users have reported zero maintenance over several years.
Download Oil in water product range brochure
GENERAL |
||||
Measurement |
Oil in water |
|||
Measurement principle |
Laser Induced UV fluorescence |
|||
Range |
0 to 20,000 ppm |
|||
User may select a range from 0 to 10 ppb to 0 to 20,000 ppm |
||||
Cleaning |
Automatic ultrasonic |
|||
Accuracy |
± 1 % of measurement range |
|||
Repeatability |
> 99 % |
|||
Response time |
< 1 second, continuous results |
|||
Process temp. range |
0 to 100 °C (option to 180 °C) |
|||
Process pressure range |
0 to 35 barg (180 barg optional) |
|||
Process Flow Rate |
up to 10 m/s |
|||
Ambient temp. range |
-20 to 55 °C |
|||
Display |
Interactive touch screen |
|||
Hot insertion/extraction |
Optional |
|||
Conduit length |
3 m to 50 m |
|||
Spectrometer Specification (1000 model only) |
||||
Emission wavelength range |
400 to 1,100 nm |
|||
Resolution |
0.5 nm |
|||
Certification |
||||
European |
CE marked |
|||
Ingress protection |
IP66 Enclosure, IP68 Probe |
|||
ATEX/IECEx |
Exd II 2G IIB T4 |
|||
CSA |
Class 1 Div 1 |
|||
Shipping |
||||
Shipping weight |
Approx. 85 Kg including stand |
|||
MECHANICAL |
||||
Dimension |
1120 H x 600 W x 640 D mm (optional variants) |
|||
Clear space |
500 mm front and rear |
|||
Material (wetted parts) |
316 SS (option of hastalloy, inconel, CR25, CR 22, titanium, monel) |
|||
ATEX box |
Aluminium (316SS optional) |
|||
Pressure to cell |
0 to 35 BarG (design pressure 180 BarG) |
|||
Process connection |
2" ANSI flange (options available) |
|||
ELECTRICAL / UTILITIES |
||||
Power voltage |
110 or 230 VAC |
|||
Power frequency |
50 or 60 Hz |
|||
Power consumption |
60 W normal, 300 W peak |
|||
Instrument air |
Not required |
|||
Outputs |
||||
Current |
4 to 20 mA passive (modbus over HART optional) |
|||
Alarm relay 1 |
Volt free relay configurable to system alarm or ppm setpoint |
|||
Alarm relay 2 |
Volt free relay configurable to system alarm or ppm setpoint |
|||
Digital 1 |
Ethernet |
|||
Digital 2 |
2 wire ADSL |
|||
Wireless |
Wi-fi (optional) |
|||
Remote access |
VNC, master remote manager |
|||
Internal data storage |
> 10 years |
|||
All specifications subject to change |
VIDEO:
Oil in water analysis - ultrasonic cleaning demonstration video
Oil in water technology overview
DRAWINGS: