Volumetric water in oil system

Aquamax KF Volumetric

easy to use
ppm, mg/l, %
2 built-in pumps
programmable

Download a datasheetRequest a quoteCall us

 

A volumetric titration system

This easy to use titration system has an intuitive keypad with a function guide display. It has an automatic system to eliminate bubbles which can form in the liquid circuit.

Two 5 ml syringes are provided, one for the KF reagent, and the other for automatic dosing of the standard concentration. Two pumps are also provided for dispensing the solvent and for emptying the vessel.

The results can be expressed in terms of ppm, mg/l or %, and can be connected to an external computer keyboard for text and sample information data entry. The Aquamax KF Volumetric connects to different types of computers, printers, and balances, and automatically stores data for the last 55 analysis results.

The optional TiCom software allows viewing and printing of all calculation input, calibration, and titration measurement data, allowing reports on multiple samples to be generated directly and exported to Microsoft Excel.

TiCom software

GENERAL

Measurement

Water content of oil samples

Measurement principle

Volumetric Karl Fischer Titration

Configuration

Benchtop

Measuring ranges

From 0.1 mg up to 100% water

Dispensing Accuracy

≤0.2 % for volumes higher than 10 % of syringe

Dispensing Reproducability

±0.1 % for volumes higher than 10 % of syringe

Resolution

1/40,000 of syringe volume

Syringe volume

Standard syringe 5 ml

Operating temperature

+15 to +40 °C (≤80 % relative humidity, non condensing)

Storage temperature

-10 to +50 °C (≤80 % relative humidity, non condensing)

Display

Graphic backlit LCD, 128 x 64 dots

Languages

English, Spanish, French, Italian

Keypad

Membrane, 7 keys, guaranteed up to 6 million strokes per key

Non-volatile memory

Up to 10 titration programmes, clock/calendar, header texts: 2 lines x 40 characters, name of up to 4 operators, programme comments: 8 lines x 40 characters, up to 55 results stored in the data logger

Certification

European

CE marked

Electrical

EC, EN 61010

EMC

EC, EN 50081-2 and EN 50082-2

Shipping

Shipping weight

4 Kg approx.

MECHANICAL

Dimension

130 x 160 x 300 mm

Enclosure material

ABS and enamelled steel

Syringe material

borosilicate glass and PTFE

Electrovalve material

PTFE and KEL-F

Tubes material

PTFE

ELECTRICAL / UTILITIES

Power

90 to 264 VAC, 47 to 63 Hz, 24 VDC

Polarised electrode

BNC connector

External keyboard

miniDIN connector

 

RCA connector

Outputs

Digital 1

RS232C bidirectional for PC or printer. RS232C for balance, telephone and stirrer control

All specifications subject to change