High precision chilled mirror device for dew point measurement
MBW 373
laboratory reference
high precision
fast response
A high precision chilled mirror device to measure the dew point, or frost point, of a gas.
The temperature of a polished mirror is closely controlled so that a layer of condensation (dew or frost) is maintained at a constant thickness. The 373 chilled mirror series from MBW offers a wide range of instruments suitable for precision measurements of humidity.
The mirror is illuminated, and a light-sensitive detector determines the thickness of condensation on the mirror surface. Automatic control of the mirror temperature, using thermo-electric coolers (or a peltiers device), cools the mirror to a temperature at which condensation occurs. Once a thin layer of condensation is present, careful control of the mirror temperature maintains a constant thickness of condensation.
Quick response at low level measurements can be achieved with ORIS, a unique facility to improve response time without affecting accuracy or stability. Download the data sheet for the MBW373 for more information
S |
H |
HX |
L |
LX |
LHX |
|
GENERAL |
||||||
Range: Dew/Frost point °C |
-60 to +20 |
-40 to +70 |
-40 to +95 |
-75 to +20 |
-95 to +20 |
-75 to +95 |
Thermoelectric mirror cooling |
3-stage |
|||||
ORIS |
Optional |
No |
Yes |
|||
Cooling |
Air/Water |
Refrigeration |
||||
Measuring head heating °C |
No |
+80 |
+100 |
+40 |
+40 |
+100 |
Internal pump |
Optional |
Yes |
Optional |
Yes |
||
Accuracy |
± 0.1 °C Dew/Frost Point |
|||||
Reproducability |
± 0.05 °C Dew/Frost Point |
|||||
Display |
Active Matrix Color Graphic LCD |
|||||
Digital I/O |
RS-232 |
|||||
Analogue output |
2 selectable ± 10 V 4 to 20 mA |
|||||
Gas couplings |
Swagelok® 6 mm or 1/4" VCR Cajon® 1/4" optional |
|||||
Mirror temperature sensors |
RTD (Pt-100) |
|||||
External temperature sensors |
RTD (Pt-100) measuring range: -50 to +100 °C |
|||||
Sample gas pressure |
2000 hPa abs. 1 MPa abs. optional |
|||||
Sample gas flow rate |
0 to 2 l/min |
|||||
Sample gas circuit |
Stainless steel, 316 / 316L, electro polished |
|||||
Operation temperature |
+15 to +35 °C |
|||||
Storage temperature |
-10 to +50 °C |
|||||
Ambient humidity |
90% RH max., non-condensing |
|||||
MECHANICAL |
||||||
Width |
449 mm |
520 mm |
449 mm |
|||
Height |
236 mm |
255 mm |
236 mm |
|||
Depth |
461 mm |
511 mm |
500 mm |
511 mm |
||
Weight |
20 kg |
25 kg |
25 kg |
35 kg |
45 kg |
38 kg |
ELECTRICAL |
||||||
Voltage |
100-120 VAC / 200-240 VAC, 50/60 Hz (auto-switching) |
100-120 VAC, 50/60 Hz or 200-240 VAC, 50/60 Hz |
||||
Power consumption |
200 W |
300 W |
300 W |
300 W |
500 W |
500 W |
All specifications subject to change |
MANUALS:
Chilled Mirror 373 Operations Manual
CALCULATORS AND CONVERSION CHARTS:
TECHNICAL BULLETINS:
Explanation of relative humidity
DATASHEET: