A precision dew point measurement with remote heads
MBW 473
cable mounted measuring heads
climate chamber validation
ForcefrostTM function
Flexible measurement options with the 473 chilled mirror.
The determination of dew point by monitoring condensation on a temperature controlled mirror allows a highly accurate (±0.1 °C) measurement. With a typical service life of more than 15 years, MBW chilled mirror hygrometers provide highly accurate and stable measurements over a long service life.
The 473 is a transfer reference standard instrument designed to validate a wide variety of climate chambers and moisture generator systems.
Optimum Response Injection System (ORIS)
In common with other models in the range, the MBW473 has the ORIS facility, reducing stabilisation time when measuring at low dew point levels. ORIS uses a pre-programed vapour injection procedure to accelerate the formation of a frost layer on the chilled mirror surface.
ForcefrostTM function
Below 0 °C water may condense as either liquid or solid. This is a particular problem when working at dew points just below 0 °C, and can cause up to a 3 °C difference in condensation temperature. For correct calculation of relative humidity (rh), the phase of the condensate must be known. When the Forcefrost function is activated at a user defined temperature the mirror over-cools to ensure that a frost is formed when below 0 °C, and to remove the uncertainty.
COMPATIBLE MEASURING HEADS |
||
|
RP2 Measuring HeadTo allow optimum placement in working volumes of air, the RP2 head includes a 2-stage thermoelectric cooler (Peltier) element in a compact probe. Range: -20 to +70 °C Order code: 103692 |
|
SH2 Measuring HeadIdeal for climate chamber validation, humidity generators, engine test cells and on-site calibration projects, this flow-through head assembly includes a variable speed fan that pulls a consistent airflow across the mirror. With the fan removed, the SH2 can be connected to applications using tubing and a standard 6 mm or 1/4 " fitting. Range: -30 to +70 °C Order code: 103693 |
||
Precision Temperature MeasurementBoth the RP2 and the SH2 versions of the 473 are supplied with a 4-wire PT1000 Platinum Resistance thermometer (PRT). The temperature probe can be directly connected to the measuring head, or by cable to the 473 back panel, allowing precise temperature measurements to be made for the calculation of rh. Range: Head Mounted: -20 to +80 °C Cable mounted: -50 to 100 °C RP2: Ø3 x 30 mm PRT, 0.5 m cable SH2:Ø2 x 100 mm PRT, 0.5 and 3 m cables |
||
Transit CaseEach 473 is supplied with a robust IP65 transit case for safe transportation to site for calibration projects. |
Order Code |
Description |
103693 |
Model 473, with SH2 Head assembly on 2 m cable, range -30 to +70°C dew point, 100x 2mm PRT temperature probe with 0.5 and 3m cable. Complete with transport case |
103692 |
Model 473, with RP2 Head assembly on 2 m cable, range -20 to +70°C dew point, 30 x 3mm PRT temperature probe with 0.5 cable. Complete with transport case |
Options |
|
103846 |
473 upgrade factory calibration to ISO17025 |
102662 |
Analogue outputs, user programmable, 2 x 4to 20mA or 0 to 10V |
100282 |
0.1% accuracy internal barometric pressure sensor |
103954 |
0.01% accuracy internal barometric pressure sensor |
103632 |
Additional 1 year warranty upgrade (maximum 3 years) |
For a complete range of options and accessories, please contact us |
RP2 |
SH2 |
|||||
GENERAL |
||||||
Measuring range |
||||||
Frost/Dew point |
-20 to +70 °C |
-30 to +70 °C |
||||
Relative humidity |
+5 to +100 %rh |
+2 to +100 %rh |
||||
Temperature (head mount PRT) |
-20 to +80 °C |
-20 to +80 °C |
||||
Temperature (cable mount PRT) |
-50 to +100 °C |
-50 to +100 °C |
||||
Accuracy |
||||||
Frost/Dew point |
≤ ± 0.1 °C (-20 to +70 °C), ± 0.2 °C |
|||||
Temperature |
≤ ± 0.1 °C |
|||||
Reproducability |
||||||
Frost/Dew Point |
≤ ± 0.05 °C |
|||||
Temperature |
≤ ± 0.05 °C |
|||||
Standard Features |
||||||
Temperature probe |
Ø3 x 30 mm PRT, 0.5 m cable |
Ø2 x 100 mm PRT, 0.5 and 3 m cables |
||||
Digital I/O |
RS-232 |
|||||
Display |
5.7" LCD with colour touch screen |
|||||
Thermoelectric mirror cooling |
2-stage with typically 50 °C depression at 20 °C ambient |
|||||
Mirror temperature sensors |
Platinum resistance thermometer (Pt100) |
|||||
Gas connections |
6 mm or 1/4 " Swagelok (SH2 only) |
|||||
Transport case |
Custom fit foam lined Peli 1550 |
|||||
Power cable |
2.5 m |
|||||
Operating instructions |
English |
|||||
Calibration certificate |
Factory calibration: 5 points FP/DP, 3 points temperature. Upgrade to SCS accredited ISO17025 calibration available |
|||||
Optional Features |
||||||
Internal barometric pressure sensor |
0.1 % or 0.01 % accuracy, 700 to 1200 mbar |
|||||
Analogue outputs |
User programmable, 2x 4 to 20 mA or 0 to 10 V |
|||||
Additional Information |
||||||
Instrument operating conditions |
0 to 40 °C, 90 %rh non-condensing |
|||||
Measuring head operating conditions |
-20 to +80 °C, 99 %rh non-condensing |
|||||
Storage conditions |
-20 to +50 °C |
|||||
MECHANICAL |
||||||
Instrument |
In Transport Case |
|||||
Dimensions |
W310 x H155 x D265 mm |
W510 x H220 x D450 mm |
||||
Weight |
5 kg |
12 kg |
||||
Protection |
IP54 |
IP65 |
||||
ELECTRICAL |
||||||
Power supply |
100 to 120 VAC / 200 to 240 VAC, 50/60 Hz, 100 Watt (auto switching) |
|||||
All specifications subject to change |
MANUALS:
Chilled Mirror 473 Operations Manual
CALCULATORS AND CONVERSION CHARTS:
TECHNICAL BULLETINS:
Explanation of relative humidity
DATASHEET: